X-Ray and Gamma Personal Dosimeters PM1610/PM1610-01/PM1610A/PM1610A-01

Professional personal dosimeter for measuring continuous and pulse X-Ray and Gamma Radiation in the energy range 20 keV – 10 MeV. The dosimeters are recommended to protect professionals, who work with or around X-ray or gamma radioactive materials. The possible use areas include but not limited to: hospitals, mines, customs officers, nuclear stations personnel etc.
From Jan 2013 only grey color is available in stock.
where to buy

Polimaster’s miniature professional personal dosimeter designed to monitor and measure the personal dose equivalent Hp(10) and personal dose equivalent rate Ĥp(10) from both gamma and X-ray radiation. It is capable to measure both continuous and pulse (from 1 msec) X-ray and gamma radiation in the energy range from 20 keV to 10 MeV.

The dosimeter allows for setting up independent threshold alarm levels for dose and dose rate (two for each). When those thresholds are exceeded, the instrument alerts user with audio, visual and vibration alarms. Audio and vibration alarms can be enabled/disabled in the instrument setting menu.

The intrument’s non-volatile memory allows for manual or automatic recording up to 7500 events of the dose rate change, acute dose levels, time and levels when the present threshold values were exceeded.

Key features of personal radiation dosimeters:

  • Measures continuous and pulse X-Ray and gamma radiation;
  • Alerts user via audio, visual and vibration alarms when preset thresholds are exceeded;
  • Records manually or automatically for up to 7500 events in the instruments' non-volatile memory;
  • Communicates with a PC via USB channel with the simultaneous battery charging;
  • Use of different types of wireless readers in compliance with ISO15693 (for PM1610-01 and PM1610A-01);
  • Low battery warning;
  • User friendly interface with two button operation.


Instrument’s shockproof rubberized case and the LCD with fluorescent backlight allows for easy operation even in the harshest, most unfavorable environments or weather conditions. The instrument is recommended for use by hospitals, mines, customs officers, nuclear stations personnel, and any professional who work with or around X-ray or gamma radioactive material.

Attention!

Instrument’s battery will gradually run down during a long-term storage (even if the instrument is switched off). Over-discharged batteries could lose some capacity. For this reason Polimaster recommends users to charge the battery periodically (at least once per six months).

Personal Dose Tracker software, developed by the Polimaster company, enables serial connection up to 100 units of PM1610 X-Ray and gamma radiation personal dosimeters. Software enables information transfer by USB-protocol to a personal computer to analyze, process, and generate corresponding databases in a control center or an expert center.

Standards compliance

Designed to meet ANSI N42.20 - 2006 and IEC 61526: 2010 requirements

 

Back to home page

 

View other dosimeters

 

Applications:


PM1610PM1610-01PM1610APM1610A-01
Detector
GM tubeGM tubeGM tubeGM tube
DER Measurement Range
0.1 µSv/h to 10.0 Sv/h0.1 µSv/h to 10.0 Sv/h0.1 µSv/h to 10.0 Sv/h0.1 µSv/h to 10.0 Sv/h
Dose Rate Accuracy
±(15+0.0015/H)%
(in range 0.1 µSv/h - 10 Sv/h,
where H - dose rate in mSv/h)
±(15+0.0015/H)%
(in range 0.1 µSv/h - 10 Sv/h,
where H - dose rate in mSv/h)
±(10+K1/H+K2H) %,
where H - dose rate in mSv/h
K1 - coefficient 0.0015 mSv/h
K2 - coefficient 0.0015 (mSv/h)-1
±(10+K1/H+K2H) %,
where H - dose rate in mSv/h
K1 - coefficient 0.0015 mSv/h
K2 - coefficient 0.0015 (mSv/h)-1
DE Measurement Range
continuous photon radiation:
0.05 µSv to 10.0 Sv

pulsed photon radiation:
10 µSv to 10.0 Sv
of continuous photon radiation
0.05 µSv - 10 Sv
(5 µR - 1000 R)
of pulsed photon radiation
(pulse duration not less than 1 ms)
10 µSv - 10 Sv
(1 mR - 1000 R)
of continuous photon radiation
0.05 µSv - 20 Sv
(5 µR - 2000 R)
of pulsed photon radiation
(pulse duration not less than 1 ms)
10 µSv - 20 Sv
(1 mR - 2000 R)
of continuous photon radiation
0.05 µSv - 20 Sv
(5 µR - 2000 R)
of pulsed photon radiation
(pulse duration not less than 1 ms)
10 µSv - 20 Sv
(1 mR - 2000 R)
Dose Accuracy
± 20%± 20%± 20%± 20%
Energy range
0.020 to 10.0 MeV0.020 to 10.0 MeV0.020 to 10.0 MeV0.020 to 10.0 MeV
Energy response relative to 0.662 MeV
(137Cs)
-60% from 20 keV to 33 keV
-40% from 33 keV to 48 keV
±30% from 48 keV to 3 MeV
±50% from 3 MeV to 10 MeV
-60% from 20 keV to 33 keV
-40% from 33 keV to 48 keV
±30% from 48 keV to 3 MeV
±50% from 3 MeV to 10 MeV
-60% from 20 keV to 33 keV
-40% from 33 keV to 48 keV
±30% from 48 keV to 3 MeV
±50% from 3 MeV to 10 MeV
-60% from 20 keV to 33 keV
-40% from 33 keV to 48 keV
±30% from 48 keV to 3 MeV
±50% from 3 MeV to 10 MeV
Survive after momentary influence of maximum
permissible gamma radiation within 5 min:

100 Sv/h (10000 R/h)100 Sv/h (10000 R/h)100 Sv/h (10000 R/h)100 Sv/h (10000 R/h)
Ingress protection rating
IP65IP65IP65IP65
Drop test
1.5 m (4.9 ft) onto hard wooden surface1.5 m (4.9 ft) onto hard wooden surface1.5 m (4.9 ft) onto hard wooden surface1.5 m (4.9 ft) onto hard wooden surface
Water resistant
xxxx
Data channel
Alarm type
visual, audio, vibrationvisual, audio, vibrationvisual, audio, vibrationvisual, audio, vibration
Additional
Countdown timer
xxxx
Display of time to critical exposure
xxxx
Functions
xxxx
Calendar mode
xxxx
PC CommunicationUSBUSB, ISO15693USBUSB, ISO15693
Power
Batteries
built-in rechargeable battery (charged via USB)built-in rechargeable battery (charged via USB)built-in rechargeable battery (charged via USB)built-in rechargeable battery (charged via USB)
Battery discharge control
LCD indicationLCD indicationLCD indicationLCD indication
Battery lifetime (average DER not more than 0.3 µSv/h,
alarms activated for not more than 20 s/24 hours)
30 days30 days30 days30 days
Physical Parameters
Dimensions
58 х 59 х 20 mm
(2 2/7'' x 2 2/7'' x 5/7'')
58 х 59 х 60 mm
(2 2/7'' x 2 2/7'' x 5/7'')
58 х 59 х 20 mm
(2 2/7'' x 2 2/7'' x 5/7'')
58 х 59 х 20 mm
(2 2/7'' x 2 2/7'' x 5/7'')
Weight
70 g (2.5 oz)70 g (2.5 oz)70 g (2.5 oz)70 g (2.5 oz)
Environmental Characteristics
Temperature
-20 up to +50 °С
(from -4 up to +122 °F)
-20 up to +50 °С
(from -4 up to +122 °F)
-20 up to +50 °С
(from -4 up to +122 °F)
-20 up to +50 °С
(from -4 up to +122 °F)
Humidity
up to 98% at 35°C (95°F)up to 98% at 35°C (95°F)up to 98% at 35°C (95°F)up to 98% at 35°C (95°F)